Modular PMD/PDL Measurement Systems
 
 
Related Products


Two Basic Systems to Choose From

  • Fast In-Line PMD/PDL Analysis System for Fibers: PMD5000FIN
  • High Dynamic Range PMD/PDL Analysis System for Narrow Bandwidth Components: PMD5000HDR

Features of the Modular PMD/PDL Analysis System PMD5000

  • Measurements Based on Jones-Matrix Eigenanalysis (JME)
  • 2 Polarimeter Versions: Rotating Waveplate or Fiber Based
  • Bandwidth from 1520 to 1630nm (others on request)
  • Modular and Flexible
  • Excellent Accuracy
  • Supports External Laser Sources*

Applications for the PMD5000

  • PMD Measurements According ITU-T G.650
  • Reference Method in the Range 1520 - 1630nm
  • Wavelength Dependent PDL, SOP, DOP Analysis
  • PMD/PDL Analysis of Narrow Bandwidth Components
  • Installed Fiber JME Analysis
  • Field Applications

Introduction
The PMD5000 is a modular system for Jones Matrix Eigenanalysis based PMD/PDL measurements built onto the TXP platform. It is a powerful and flexible solution for all kinds of PMD/PDL related measurement tasks in laboratories as well as field applications including the analysis of installed fibers.

Modular Design
Depending on the application the PMD5000 Systemcomprises different modules that can be used independently as well. The PMD5000 integrates a powerful tunable laser source but supports external sources as well.*

Analysis of Narrow Bandwidth Components
The software package version 2 for the Thorlabs PMD/PDL Analysis System PMD5000 series enhances its application especially for the analysis of narrow bandwidth components. The PMD5000 system, which is based on Jones Matrix Eigenanalysis, now allows the analyzing parameters to be changed after the device under test (DUT) has been measured. A single data set of the DUT can be analyzed for different wavelength ranges or at different zoom levels. This makes repetitive measurements with optimized parameters obsolete and contributes to speed and flexibility of the analysis. For more information please see the "User Interface" tab.

Besides the optional fast tunable laser source, ECL5000D, the PMD5000 consists of the Deterministic Polarization Controller, DPC5500, and a polarimeter module - either the fast in-line polarimeter IPM5300 or the High Dynamic Range Terminating Polarimeter, PAX5720IR3 - plus a specific software package providing powerful PMD/PDL analysis routines. A test version of this is integrated into the TXP application software which can be found at the TXP Software tab on the TXP platform page. Please see the Components tab for the modules of the 4 standard PMD5000 Systems.

For further information, please contact our Tech Support Team.


*The external laser source option has reduced measurement speed.

Components of the PMD5000 Standard Systems

Systems with Built in Tunable Laser Source (TLS):

Fast Inline Measurements: PMD5000FIN-1

  • TXP5016: 16 Slot Mainframe
  • ECL5000D: Tunable Laser Source 1519 — 1632 nm
  • DPC5500: Deterministic Polarization Controller
  • IPM5300: High Speed In-Line Polarimeter

High Dynamic Range Measurements: PMD5000HDR-1

  • TXP5016: 16 Slot Mainframe
  • ECL5000D: Tunable Laser Source 1519 — 1632 nm
  • DPC5500: Deterministic Polarization Controller
  • PAX5720IR3: High Dynamic Range Terminating Polarimeter

Systems for External Tunable Laser Sources:

Fast Inline Measurements: PMD5000FIN-2

  • TXP5016: 4 Slot Mainframe
  • DPC5500: Deterministic Polarization Controller
  • IPM5300: High Speed In-Line Polarimeter

High Dynamic Range Measurements: PMD5000HDR-2

  • TXP5016: 4 Slot Mainframe
  • DPC5500: Deterministic Polarization Controller
  • PAX5720IR3: High Dynamic Range Terminating Polarimeter

All systems come with a preconfigured laptop for plug and play.

Specifications

ModelsPMD5000FIN-1PMD5000FIN-2PMD5000HDR-1PMD5000HDR-2
Optical Parameters
Wavelength Range1520 - 1630nm*
Dynamic Range45dB60dB
Measured ValuesDGD, PDL, Mean and RMS Values of PMD, Plus 2nd Order PMD, PMD Coefficient
Polarimeter TechnologyFiber BasedRotating Waveplate
Measurement Rate, Max 0.3s per StepDepends on TLS 0.3s per stepDepends on TLS
DGD Measurement Range0.001 to 400ps
DGD Measurement Reproducibility0.01ps
PDL Measurement Range30dB50dB
PDL Measurement Reproducibility< 0.02dB
General Technical Data
Operating Temperature Range5 to 40°C
Storage Temperature Range-40 to 70°C
Optical Input / Output ConnectorFC / APCFC / APC (Input of Polarimeter FC / PC )
Warm up time for Rated Accuracy< 15min
Mains Voltage100V / 240V ± 10%
Mains Frequency50Hz / 60Hz ± 5%
Maximum Power Consumption400VA
Dimensions (W x H x D)449 x 133 x435mm³
Chassis19", 3U
Weight10.5kg8.5kg10.5kg8.5kg

*Others available Upon request
All data are valid at 23 ±5°C and 45 ±15% relative humidity)

For further information about the TXP platform, please see the TXP platform page.

PMD5000 Series User Interface

The PMD5000 System offers a wide range of measurement data. The insertion loss (IL) and the polarization dependent loss (PDL) are displayed in a single graph. The differential group delay (DGD) versus wavelength is shown in another graph with an additional histogram. The phase difference between two Jones matrices is given in a third diagram. The second order PMD and therefore the second order DGD versus wavelength can also be measured and is displayed separately with a histogram. Another measurement is the principle state of polarization shown in a diagram in the form of Stokes parameters. All data are also available as numerical values.

The following figure shows the GUI of the measurement mode with the DGD graph.

GUI Screen Shot

The green vertical lines show the wavelength range for which the analysed parameters are calculated (upper left "Calculations" pane). The blue vertical line shows a specific single wavelength (cursor) for which the parameters are calculated (lower left "Cursor" pane).

Both the wavelength range and the cursor wavelength can be set and changed after the actual measurement has been performed. This allows quick finetuning of the wavelength range of interest for the analysis without running a measument again.

Our Tech Support Department is available to assist you in customizing the configuration for your specific application. Please choose your nearest contact partner.

Standard Applications Overview
Standard PMD and PDL Analysis System for Fibers
(PMD5000FIN-1)
Standard PMD and PDL Analysis System of Narrow Bandwidth Devices
(PMD5000HDR-1)
Standard PMD and PDL Analysis System with External Laser Source
(PMD5000FIN-2 for Fiber Analysis, PMD5000HDR-2 for Component Analysis)
Splitted PMD and PDL Analysis System for Installed Fibers
(Configuration Available Upon Request)
PMD and PDL Analysis System for Optical Networks
(Single Transmitter and Several Receivers)

(Configuration Available Upon Request)



PMD and PDL Analysis System for a Live Fiber with Traffic
(Configuration Available Upon Request)


PartPMD5000FIN-1PMD5000FIN-2PMD5000HDR-1PMD5000HDR-2
TXP5016 16 Slot Mainframe
(TXP5016)
xxxx
ECL5000D Tunable Laser Source 1519 to 1632nm
(ECL5000D)
x x 
DPC5500 Deterministic Polarization Controller
(DPC5500)
xxxx
IPM5300 High Speed In-Line Polarimeter
(IPM5300)
xx  
PAX5720IR3 Terminating Polarimeter
(PAX5720IR3)
  xx
Preconfigured Laptop with Software Installedxxxx
FC/APC Patchcord 30cmx x 
Ethernet Cable 1.5mxxxx
Software CD-ROMxxxx
LabVIEW™ and LabWINDOWS™/CVI Driver Setxxxx
Operating Manualxxxx
Loading Support DocumentsLoading...

Please Give Us Your Feedback
 
Email   Feedback On
(Optional)
Contact Me:
Your email address will NOT be displayed.
 
 
Please type the following key into the field to submit this form:
Click Here if you can not read the security code.
This code is to prevent automated spamming of our site
Thank you for your understanding.
  
 
Would this product be useful to you?       Little Use   1 2 3 4Very Useful

Enter Comments Below:
 
Characters remaining 8000    
Posted Comments:
No Comments Posted
Loading Tag cloudLoading...


Order
Based on your currency / country selection, your order will ship from Newton, New Jersey  
  +1 QTY Part Number - Imperial Price Available/Ships
PMD5000FIN-2 Support Documentation PMD5000FIN-2 -    PMD/PDL Analyzer for external Tunable Laser and IPM5300 Polarimeter $34,260.00
  Lead Time
 
PMD5000HDR-2 Support Documentation PMD5000HDR-2 -    PMD/PDL Analyzer for external Tunable Laser and PAX5720IR3 Polarimeter $30,660.00
  Lead Time
 
PMD5000FIN-1 Support Documentation PMD5000FIN-1 -    PMD/PDL Analyzer with internal Tunable Laser and IPM5300 Polarimeter $59,640.00
  Lead Time
 
PMD5000HDR-1 Support Documentation PMD5000HDR-1 -    PMD/PDL Analyzer with internal Tunable Laser and PAX5720IR3 Polarimeter $56,040.00
  Lead Time
 

Author: lmorgus Last Updated: Mar 30, 2009
Log in  |  My Account  |  Contact Us  |  Privacy Policy  |  Home  |  Site Index
Copyright © 1999 - 2010 Thorlabs
   HACKER SAFE certified sites prevent over 99.9% of hacker crime.
Sales: 1-973-579-7227     |     Support: 1-973-300-3000